Rivertek Medical Systems

RSIM-1500 Clinical Scenarios


Using the built-in BASIC-like programming language, with any text editor you have high-resolution control of the timing and modification of any of the more than one hundred rates, conduction states, thresholds, timers, display messages, and events.


The example below shows a simple scenario.  Upon launch, it waits for an RV event and fires an RV PVC 350 milliseconds later.

10 CLR_RV_EVENT
20 IF (RV_EVENT) GOTO 35

30 GOTO 20   REM Synchronizes with R-wave

35  Display "RV Event detected"

40 CLR_RV_EVENT
50 SET_TIMER1 (350)

60 IF (TIMER1 = 0) GOTO 80

70 GOTO 60 REM Wait 350
mS
80 RPVC

90 END 


Assistance is available at no charge from Rivertek to help create and debug your scenarios.